Step 9. Press the EDIT soft key. Use the Up/down arrow key to move the marker over
one of the spurs. Press the ENTER key to set the marker.
Step 10. Compare the value of the marker to the specified allowable level of out-of-band
spurious emissions for the corresponding channel transmit frequency.
Step 11. Repeat Steps 8-11 for the remaining spurs. Use either Marker 1 again, or choose
another marker.
Figure 5-16 shows a simulated out-of-band spurious signal at 21.052 MHz from the carrier.
5-24
Chapter 5 Spectrum Analyzer Measurements
Figure 5-16. Simulated Out-of-Band Spurious Emission Measurement