3-3 Cable & Antenna Analyzer Quick Start Guide
3-6 PN: 10580-00252 Rev. A Site Master User Guide
Calibrate with OSL Calibration
1. Press the Freq/Dist main menu key and enter the appropriate frequency range
2. Press Shift then Calibrate (2) key.
3. Select Standard or FlexCal.
4. Press Start Cal and follow instructions on screen.
5. Connect Open to RF Out and press the Enter key.
6. Connect Short to RF Out and press the Enter key.
7. Connect Load to RF Out and press the Enter key.
8. Verify that the calibration has been properly performed by checking that the Cal Status
message is now displaying “ON, Standard” or “ON, FlexCal”.
Note
Refer to the Cable & Antenna Measurement Guide listed in Appendix A for
calibration details.
Figure 3-5. Calibration Setup with OSL Cal